1.General Introduction

Yokogawa SC450G is a four wire conductivity analyzer converter to measure specific conductivity with contacting electrodes, accepts sensors with cell constants ranging from 0.005 till 50/cm; 2-electrode sensors as well as 4-electrode sensors; 5 different temperature compensating elements for accurate temperature compensation. It owns Unique touch screen operation with menu structure in 5 languages, enhanced diagnostics, process trending graphics and on-screen logbook for data storage, Two mA outputs and four SPDT relay contacts with display indicators and Hart communications optional.

2.General Specifications

Items

Description

Input specification

Two or four electrode   measurement with square wave excitation, using max. 60 m cable and cell   constants from 0.005 to 50.0 cm-1

Detection method

Frequency, read-pulse   position and reference voltage are dynamically optimized.

Input ranges

Conductivity:   0.000μS/cm to 2000mS/cm, Minimum: 1μS/cm (under range 0.000μSx C); Maximum:   200mS/cm (over range 2000mS x C).
  Resistivity: 0.0 Ω x cm to 1000 MΩ x cm, Minimum: 5 Ω x cm (under range 0   Ω/C); Maximum: 1 MΩ x cm (over range 1000 MΩ/C).
  Temperature:  Pt1000: -20 to 250°C (0 to 500°F), Pt100 and Ni100: -20 to   200°C (0 to 400°F); NTC 8k55: -10 to 120°C (10 to 250°F); Pb36 NTC: -20 to   120°C (0 to 250°F)

Conductivity/Resistivity   Accuracy)

≤0.5 % of reading

Temperature Accuracy

≤0.3°C (≤ 0.4°C for   PT100)

Step Response

≤4 sec for 90 % (for a   2 decade step)

Transmission Signals

Two isolated outputs of   4 to 20mA DC with common negative. Maximum load 600 Ω. Bi-directional HART   digital communication, superimposed on mA1 (4 to 20mA) signal. Burn up   (21.0mA) or Burn down (3.6mA) to signal failure accorded with NAMUR NE43.

Temperature   compensation

Automatic or manual,   for temperature ranges.
  Reference temp.: programmable from 0 to 100°C or 30 to 210°F (default 25°C)

Compensation algorithm

According IEC 60746-3   NaCl tables (default).
  Two independent user programmable temperature coefficients, from 0 to 3.5   %/°C (°F) by adjustment or calibration.
  Matrix compensation: with conductivity function of concentration and   temperature. Choice out of 13 preprogrammed matrixes.

Display

Graphical Quarter VGA   (320 x 240 pixels) LCD with LED backlight and touch screen.

Contact Outputs

Four SPDT relay   contacts with display indicators.
  Contact outputs configurable for hysteresis and delay time.

Contact Input

Remote range switching   to 10 times the programmed range

Power Supply

85 to 265 V AC ±10 %,   47 to 63 Hz, maximum consumption 10 VA, 9.16 to 30 V DC ±15 %, maximum   consumption 10 W

 

3.Model Selection

Main Code

Suffix Code

Description

SC450G




Conductivity/Resistivity   Converter

Power

-A



AC version (100   to 240 VAC)

-U



DC version (12   to 24 VDC)

Type


-A


General purpose   version


-U


FM version

Optional codes



/UM

Universal mounting kit (panel,pipe, wall)



/U

Pipe and wall   mounting hardware



/PM

Panel mounting   hardware



/H5

Awning hood   (stainless steel)



/AFTG

G1/2



/ANSI

1/2NPT



/AM20

M20



/SCT

Stainless steel   tag plate

 

4.External Dimensions